KYORITSU KEW 5204 Digital Light Meter (Lux Meter)
KYORITSU KEW 5204 Digital Light Meter is a digital light meter (Lux Meter) for measuring illuminance in buildings, offices, industrial plants, and various work areas. It supports measurement ranges from 0.0 to 199,900 Lux, making it suitable for checking lighting conditions and evaluating the efficiency of lighting systems.
The device offers four selectable measurement ranges: 199.9 / 1,999 / 19,990 and 199,900 Lux. It has a maximum resolution of 0.1 Lux and an accuracy of ±(4% rdg + 5 dgt) at the reference temperature of 23±2°C.
The KEW 5204 features a detachable and rotatable light sensor (Detachable & Rotatable Light Sensor), allowing for positioning the sensor appropriately to measure different surfaces, making it easy to read values on the device. It also comes with a large LCD screen that has backlighting for use in low-light areas.
The KEW 5204 model supports measuring illuminance from LED lamps according to manufacturer testing conditions and includes functions such as Data Hold, MAX/MIN, Zero Adjustment, and Auto Power Off. These features make it ideal for lighting system inspections, maintenance work, and environmental condition measurements in the workplace.
General Features
- Brand: KYORITSU
- Model: KEW 5204
- Type of Product: Digital Light Meter / Lux Meter
- Luminance Measurement Range: 0.0-199,900 Lux
- Selectable measurement ranges: 199.9 / 1,999 / 19,990 / 199,900 Lux
- Maximum resolution of 0.1 Lux
- Accuracy ±(4% rdg + 5 dgt) at 23±2°C
- Silicon Photodiode Light Sensor Type
- The light sensor head can be detached and rotated to adjust position
- Supports measuring illuminance from LED lamps under appropriate usage conditions
- Data Hold function for holding displayed values on the screen
- MAX/MIN function for displaying maximum and minimum values
- Zero Adjustment function for adjusting zero before measurement
- Large LCD screen with backlighting
- Auto Power Off system that automatically turns off after approximately 30 minutes of inactivity
- Batteries: AA 1.5V x 2
- Complies with IEC 61326 and JIS C 1609-1:2006 standards
Technical Specifications (Technical Specifications)
| Brand | KYORITSU |
| Model | KEW 5204 |
| Product Type | Digital Light Meter |
| Measurement Parameter | Illuminance |
| Measurement Range | 0.0-199,900 Lux |
| Measurement Ranges | 199.9 / 1,999 / 19,990 / 199,900 Lux |
| Resolution | 0.1 Lux |
| Accuracy | ±(4% rdg + 5 dgt) at 23±2°C |
| Light Detector | Silicon Photodiode |
| Light Sensor Design | Detachable and Rotatable |
| Angular Deviation at 10° | ±1.5% |
| Angular Deviation at 30° | ±3% |
| Angular Deviation at 60° | ±10% |
| Angular Deviation at 80° | ±30% |
| Relative Spectral Sensitivity | Deviation from Spectral Luminous Efficiency: 9% |
| Response Time (Auto Range) | 5 Seconds or Less |
| Response Time (Manual Range) | 2 Seconds or Less |
| Functions | Data Hold, MAX, MIN, Zero Adjustment, Backlight, Auto Power Off |
| Auto Power Off | Approximately 30 Minutes |
| Operating Temperature | 0 to 40°C |
| Operating Humidity | 80% RH or Less, Non-Condensing |
| Storage Temperature | -10 to +60°C |
| Storage Humidity | 70% RH or Less, Non-Condensing |
| Power Supply | AA 1.5V (R6/LR6) x 2 |
| Continuous Measuring Time | Approximately 80 Hours |
| Dimensions (L x W x D) | 169 x 63 x 37 mm |
| Weight | Approximately 210 g |
| Applicable Standards | IEC 61326, JIS C 1609-1:2006 |
| Bluetooth | Not Supported (Available on KEW 5204BT) |
Operating Principle
The KYORITSU KEW 5204 Light Meter uses a Silicon Photodiode sensor to detect light falling onto the receiving head, which is then converted into an electrical signal for processing and displaying the illuminance value in Lux on the LCD screen.
The receiving head can be rotated and detached from the main unit, allowing users to place the sensor head on the surface they wish to measure conveniently. This helps reduce limitations when reading values while positioning the receiving head.
The device can select either Auto Range or Manual Range measurement modes with a response time of no more than 5 seconds in Auto Range mode and no more than 2 seconds in Manual Range mode, according to the manufacturer's data.
Functions
- Illuminance Measurement: Measures illuminance in Lux units
- Auto / Manual Range: Selects automatic or manual range measurement modes
- Data Hold: Keeps the displayed value for easy reading
- MAX / MIN: Displays the highest and lowest measured values
- Zero Adjustment: Adjusts zero according to the manufacturer's instructions before starting measurements
- LCD Backlight: Helps read values on the screen easily in low-light areas
- Auto Power Off: Automatically turns off after approximately 30 minutes of non-use
Usage and Measurement
- Check the readiness of the device and battery before use.
- Check the cleanliness of the light receiving sensor head.
- Turn on the device and select either Auto Range or Manual Range measurement mode.
- Follow the zero adjustment steps as specified in the manufacturer's manual.
- Position the sensor head appropriately for the surface to be measured.
- Avoid blocking the light falling onto the receiving head with your hand or body.
- Wait until the measurement value stabilizes before reading and recording it.
- Use Data Hold or MAX/MIN functions as needed based on the task requirements.
- Measure multiple positions when assessing illuminance levels of an area.
Turn off the device and properly store the sensor head after use
Usage Examples and Guidelines
- Measure light levels inside offices and buildings
- Inspect lighting systems in industrial plants
- Measure light levels at workstations and work areas
- Check lighting systems in warehouses and storage areas
- Use for assessing the lighting environment in facilities
- Test LED lamps under appropriate measurement conditions
- Use for inspecting and maintaining building lighting systems
- Apply for checking the efficiency of lighting systems
Accessory Information as Per Manufacturer
- KYORITSU KEW 5204 Light Meter with sensor head
- CARRYING CASE MODEL 9195 (Carrying Case)
- Batteries AA 1.5V x 2 pieces
- User Manual (Instruction Manual)
Important Considerations
- The KEW 5204 model is a non-Bluetooth light meter
- If Bluetooth app connectivity is required, consider the separate KEW 5204BT model
- When measuring LED lamps with dimming or pulsing effects, consider testing conditions and manufacturer limitations
- Position the sensor head appropriately and avoid obstructing the light during measurement
- Check calibration and battery readiness before use
- For measurements requiring reference to standards or legal requirements, verify the meter's calibration status
- Verify the actual accessory set shipped by the vendor before ordering