Webster KEPLER KH-20W Series / KH-75W Series / KH-92WS (Webster Hardness Tester)
KEPLER Webster Hardness Tester is a portable metal hardness testing device that uses the indentation method to assess material hardness. It comes in three series: KH-20W Series, KH-75W Series, and KH-92WS, each suitable for different types of materials, shapes, and sizes.
The instrument is designed as a mechanical system (Mechanical) that operates manually with a direct dial display, eliminating the need for electrical power during testing. It is ideal for checking the hardness of metal materials such as aluminum alloys, brass, copper alloys, and stainless steel, requiring selection based on material type.
It supports a measurement range of 0-20 HW and has an accurate measuring range according to specifications from 5-18 HW, suitable for inspecting sheet metals, pipes, and various shaped parts under the thickness and size conditions supported by each model.
The compact design allows easy one-handed operation, making it convenient for quality control in production lines, material inspection, and field metal part testing.
Features
Brand: KEPLER
Type of Product: Webster Hardness Tester
Series: KH-20W Series, KH-75W Series, and KH-92WS
Mechanical portable metal hardness tester
Uses Indentation testing method
Mechanical dial display
Manual operation (Manual Testing)
Able to measure with one hand
Measurement range: 0-20 HW
Accurate measurement range according to specifications: 5-18 HW
Precision of 0.5 HW as per catalog data
Resolution of 0.1 HW or 0.5 HW depending on the model
Selectable for aluminum, brass, copper alloys, or stainless steel based on model
Suitable for sheet metal and pipe testing according to dimensional specifications of each model
Manual calibration system
Uses Standard Hardness Block for inspection and calibration
Able to change the test stylus (Stylus)
Approximate weight: 0.5 kg
Suitable for quality control and material part acceptance work
Technical Specifications
Specifications
Details
Brand
KEPLER
Product
Webster Hardness Tester
Model Series
KH-20W Series / KH-75W Series / KH-92WS
Testing Method
Manual
Working Principle
Indentation
Display
Mechanical
Hardness Measurement Range
0-20 HW
Accurate Measurement Range
5-18 HW
Accuracy
0.5 HW
Resolution
0.1 HW / 0.5 HW (Depending on Model)
Flat Workpiece Thickness Requirement
Depending on Model; Catalog Lists 0.4-6 mm, 0.6-6 mm, 0.6-8 mm and 0.6-13 mm
Note: Sub-models differ in terms of suitable materials, hardness range, gauge size, and workpiece dimensions. Therefore, it is important to check the model code before purchasing as not all models have identical specifications.
How Webster Hardness Tester Works
The Webster Hardness Tester operates on the principle of pressing a test head onto the surface of a material. The device has mechanisms that convert the reactive force from the press into hardness values displayed on the dial in HW (Webster Hardness) units.
General use involves positioning the workpiece between the pressing parts of the tester, then squeezing the handle to apply pressure to the test head against the surface of the workpiece. The hardness value is read according to the steps outlined in the manual.
The indentation method may leave small marks on the surface of the workpiece, so it's important to consider testing positions and surface condition requirements before use, especially for finished pieces with restrictions regarding marks.
Materials Suitable for Testing (Applicable Materials)
Aluminum Alloy
Brass
Copper Alloy
Stainless Steel
The four material types are specified in the catalog for this series. The correct model must be chosen to accommodate the materials and hardness range required.
Workpiece Specifications
Machines in the Webster Hardness Tester group have different thickness requirements for sheet workpieces and internal diameter specifications for tube workpieces depending on the model, as specified by the catalog's thickness ranges such as 0.4-6 mm, 0.6-6 mm, 0.6-8 mm, and 0.6-13 mm.
For tube work, there is a minimum internal diameter requirement of 6 mm or 10 mm depending on the model.
Before selecting a machine, it's important to check the thickness, shape, and gauge size to ensure they match the actual workpiece for proper positioning and accurate measurement readings.
Calibration and Verification
The device supports Manual Full Scale Calibration and uses a Standard Calibration Block to confirm measurement values according to the manufacturer's manual.
Before use, it is advisable to check the dial, mechanism, and condition of the stylus tip, as well as verify the readings with an appropriate hardness reference block for the model.
If the device reads incorrectly compared to the reference block, follow the manufacturer’s procedures for inspection and adjustment; do not make adjustments without following proper steps.
Stylus Replacement
The stylus tip is a critical component of the Webster Hardness Tester, as per the catalog it can be replaced by the user.
Replacement should involve selecting compatible parts for the model and using appropriate tools according to the manual. After replacement, verify or calibrate the device with a reference block before actual use.
Examples and Usage Guidelines (Applications)
Check hardness of aluminum alloy
Check hardness of aluminum profiles
Check hardness of aluminum sheets
Check hardness of brass parts
Check hardness of copper alloys
Check hardness of stainless steel parts according to supported models
Check hardness of metal pipes
Used in material and metal part inspection work
Used in production quality control processes
Used to check consistency of factory-made products
Initial Usage Guidelines
Ensure the model matches the type of material being tested
Check thickness and size of the workpiece according to model specifications
Inspect condition of the stylus tip and display dial
Verify readings with a Standard Calibration Block as per manufacturer’s method
Clean surface at measurement location
Position the workpiece so that the stylus touches an appropriate area
Squeeze grip handle according to manual steps
Read hardness value from Mechanical dial
Measure multiple positions as required by quality control standards
Record inspection results in job documentation
Clean and store device after use
Standard Accessories According to Catalog (Included Accessories)
Main Unit — Webster Hardness Tester
Standard Hardness Block — Standard hardness block
Spare Stylus — Spare stylus tip
Special Wrench — Special Wrench
Screwdriver — Screwdriver
Manual — User Manual
Handle Case — Storage Box or Bag
Optional Accessories (Optional Accessories)
Standard Hardness Block — Additional Standard Hardness Block
Gauge Glass — Manufacturer's Optional Item
Note: The actual set of accessories received may vary by sub-model and kit. Please confirm details with the supplier before ordering.
Considerations Before Ordering
KH-20W Series, KH-75W Series, and KH-92WS are mechanical testers, not digital testers.
The testing principle is Indentation, not Leeb Rebound or Ultrasonic Contact Impedance.
The measuring range of 0-20 HW is the full range of the instrument, while the accuracy range as specified is 5-18 HW.
Supported materials vary by sub-model code.
The thickness of sheet workpieces and pipe size must be checked before selecting an instrument.
The resolution may be either 0.1 HW or 0.5 HW, depending on the model.
The testing pressure may leave indentations on the workpiece.
Proper contact position is required for accurate measurement.
A Standard Hardness Block should be used to calibrate or confirm values.
The correct sub-model code must match the material and shape of the workpiece.
Please check the standard accessory list and warranty conditions before ordering.