KEPLER KAQ-4A Portable Multi-Gas Detector (Portable Multi-Gas Detector)
KEPLER KAQ-4A Portable Multi-Gas Detector is a portable gas detector for industrial safety applications, designed to monitor multiple gas hazards in a single unit, supporting the measurement of oxygen (O2), carbon monoxide (CO), hydrogen sulfide (H2S), and combustible gas (Combustible Gas / LEL)
The device can be used for atmosphere inspection in work areas, industrial factories, wastewater treatment systems, and areas at risk of oxygen deficiency, toxic gas accumulation, or flammable gas. It should be used in conjunction with risk assessment procedures and relevant safety measures.
The KAQ-4A model has an oxygen measurement range of 0-30% Vol, carbon monoxide measurement range of 0-1000 ppm, hydrogen sulfide measurement range of 0-100 ppm, and combustible gas measurement range of 0-100% LEL, with power supplied by a rechargeable lithium-ion battery.
The device is rated for explosion protection Ex ib IIB T3 Gb according to the catalog data, suitable for use in areas requiring atmospheric risk assessment. Before entering hazardous areas, it is necessary to check the explosion protection certificate and actual scope of use.
General Features
- Brand: KEPLER
- Product Model: KAQ-4A
- Product Type: Portable Multi-Gas Detector
- Supports measurement of 4 types of gas, namely O2, CO, H2S, and Combustible Gas (LEL)
- Oxygen (O2) measurement range: 0-30% Vol
- Carbon monoxide (CO) measurement range: 0-1000 ppm
- Hydrogen sulfide (H2S) measurement range: 0-100 ppm
- Combustible gas measurement range: 0-100% LEL
- Accuracy according to catalog data does not exceed ±5% F.S.
- Response time T90 is less than 30 seconds
- Operable in temperatures from -20 to +50°C
- Supports relative humidity below 95% RH without condensation
- Uses a rechargeable lithium-ion battery DC 3.7V with a capacity of 1800 mAh
- Charging time according to catalog data is approximately 6-8 hours
- Has explosion protection rating Ex ib IIB T3 Gb
- Suitable for gas measurement for safety monitoring in industrial work
Technical Specifications
| Brand |
KEPLER |
| Model |
KAQ-4A |
| Product Type |
Portable Multi-Gas Detector |
| Detected Gases |
O2, CO, H2S, Combustible Gas (LEL) |
| Oxygen (O2) Measuring Range |
0-30% Vol |
| O2 Resolution |
0.1% Vol |
| Carbon Monoxide (CO) Measuring Range |
0-1000 ppm |
| CO Resolution |
1 ppm |
| Hydrogen Sulfide (H2S) Measuring Range |
0-100 ppm |
| H2S Resolution |
1 ppm |
| Combustible Gas Measuring Range |
0-100% LEL |
| LEL Resolution |
1% LEL |
| Accuracy |
≤ ±5% F.S. |
| Response Time |
T90 < 30 Seconds |
| Operating Temperature |
-20 to +50°C |
| Operating Humidity |
<95% RH, Non-Condensing |
| Battery |
Rechargeable Lithium-Ion Battery |
| Battery Voltage |
DC 3.7V |
| Battery Capacity |
1800 mAh |
| Explosion Protection Marking |
Ex ib IIB T3 Gb (As Stated in Catalog) |
| Charging Time |
Approximately 6-8 Hours |
| Sampling Method |
Refer to Model-Specific Manual |
| Ingress Protection Rating |
Refer to Model-Specific Documentation |
| Dimensions / Weight |
Refer to Model-Specific Documentation |
Gas Detection Ranges
| Gas |
Measuring Range |
Resolution |
| Oxygen (O2) |
0-30% Vol |
0.1% Vol |
| Carbon Monoxide (CO) |
0-1000 ppm |
1 ppm |
| Hydrogen Sulfide (H2S) |
0-100 ppm |
1 ppm |
| Combustible Gas (LEL) |
0-100% LEL |
1% LEL |
Gas Detection Details
1. Oxygen (O2) - Oxygen
Used to measure the amount of oxygen in the atmosphere, helping to monitor abnormal low or high oxygen levels, which may pose a risk to workers, especially in confined spaces or areas where other gases replace air.
2. Carbon Monoxide (CO) - Carbon Monoxide
Used to detect carbon monoxide, a colorless and odorless toxic gas that can be produced from incomplete combustion processes, such as engines, furnaces, and various industrial processes.
3. Hydrogen Sulfide (H2S) - Hydrogen Sulfide
Used to detect hydrogen sulfide, a toxic gas that may be found in wastewater treatment systems, drainage systems, petrochemical plants, and processes involving the decomposition of organic substances. It is not advisable to use smell as a method to assess safety from this type of gas.
4. Combustible Gas (LEL) - Flammable Gas
Used to monitor flammable gas levels, displayed as a percentage of the Lower Explosive Limit (LEL), to help identify the risk of gas accumulation. The response capability depends on the type of sensor, reference gas, and calibration of the device.
Working Principle of Multi-Gas Detector
The Multi-Gas detector uses sensors to detect each type of gas and processes the signal to display the gas concentration in appropriate units, such as ppm, % Vol, and % LEL.
The measured values can be used for monitoring the atmosphere in the working area. Users must check the alarm settings and the operating status of the device according to the manual and safety requirements.
Detection of flammable gas must consider the target gas type and the calibration gas used, as the readings may vary depending on the sensor's response to each type of gas.
Preparation Before Use
- Check the model number and type of gas installed in the device to ensure they match the measurement task.
- Check the battery and operational status of the device before use
- Check the calibration schedule and perform Bump Test according to the manufacturer's manual and the employer's safety procedures
- Check that the gas inlet is not blocked or contaminated
- Ensure the alarm settings are in accordance with laws, standards, and the risk assessment of the organization
- Check the protective equipment and hazardous area requirements before using the device
- Use the gas sampling method as specified in the KAQ-4A manual
- If the device alarms or malfunctions, immediately follow the employer's emergency procedures
Examples and Usage Guidelines
- Atmospheric inspection in confined spaces according to safety procedures
- Gas detection work in petrochemical and oil industry factories
- Inspection of wastewater treatment systems and drainage pipes
- Maintenance of machinery and piping systems in industrial factories
- Inspection of areas where flammable gas accumulation may occur
- Atmospheric monitoring in steel industries, power plants, and shipyards
- Inspection of oxygen levels and toxic gases in work areas
- Used in conjunction with gas risk assessments before and during work
Safety Considerations
- The gas detector is a monitoring device and cannot be used as a substitute for permits for work in confined spaces or other safety measures
- Bump Test and calibration should be performed according to the schedule specified in the manual or the employer's requirements
- Check the sensor type and gas used for calibration, especially for flammable gas detection sensors
- Do not bring the device into hazardous areas unless the Explosion Protection certificate covers the intended use
- Avoid use in temperatures or humidity levels exceeding the specifications
- If the device alarms for hazardous gas levels, follow the emergency plan and evacuate to a safe area according to the procedure
- Do not charge the battery in hazardous areas unless the manual and certificate explicitly permit it
- Check the components, certificates, and calibration documents before ordering