K1224 IN, KEPLER, Digital High Voltage Insulation Tester, 12kV Insulation Tester, 12000V Insulation Tester, High Voltage Insulation Resistance Meter, Digital Megohmmeter, High Voltage Megger, 24TΩ Insulation Tester, 12000V Megohmmeter, PI DAR Insulation Tester, High Voltage Insulation Resistance Tester, 12kV Insulation Tester, Digital Insulation Tester, Industrial Insulation Resistance Meter
Digital High Voltage Insulation Tester KEPLER Model K1224 IN (12kV)
KEPLER K1224 IN Digital High Voltage Insulation Tester is a digital high voltage insulation tester designed for testing and evaluating the condition of electrical equipment and high-voltage power systems. It supports a maximum test voltage of 12,000VDC (12kV) and can measure up to 24TΩ insulation resistance according to specified measurement ranges and conditions.
The device features a large color display for real-time testing results, including insulation analysis functions such as IR, IR(t), DAR, PI, SV, DD, and Ramp Test. It displays and records real-time values of insulation resistance, voltage, and current, supports date and time recording of test results, and exports data via USB for further inspection and analysis.
Maximum Insulation Resistance Measurement Range: 24TΩ at test voltage of 12kV
Display: Large Color Display
Insulation Analysis Functions: IR, IR(t), DAR, PI, SV, DD and Ramp Test
Data Logging: Records test results with date and time
Data Connectivity: USB
Casing Structure: Dual Case
Power Supply: Internal Battery and AC Power Source
Test Voltage Settings
Specified Test Voltages for K1224 IN: 500V / 1000V / 2500V / 5000V / 12000V
Voltage Adjustment from 100V to 1kV: Increment or decrement by 10V each step
Voltage Adjustment from 1kV to 10kV: Increment or decrement by 25V each step
Test Voltage Accuracy: +4%, -0%, ±10V as specified in the catalog at a load of 1GΩ and temperature range of 0-30°C
Note: The catalog specifies detailed pressure adjustment up to 10kV, while the K1224 IN model has a maximum test voltage of 12kV. Therefore, it is advisable to check the pressure adjustment steps beyond 10kV directly with the user manual.
Insulation Resistance Measurement Range
Test Voltage
Measurement Range / Accuracy ±5%
Measurement Range / Accuracy ±20%
500V
0.4MΩ-100GΩ
100GΩ-1TΩ
1000V
0.8MΩ-200GΩ
200GΩ-2TΩ
2500V
2MΩ-500GΩ
500GΩ-5TΩ
5000V
4MΩ-1TΩ
1TΩ-10TΩ
10000V
8MΩ-2TΩ
2TΩ-20TΩ
12000V
9.6MΩ-2.4TΩ
2.4TΩ-24TΩ
Electrical Measurement Data
Short Circuit Current: 6mA
Voltage Measurement Range AC/DC: 30-660V
Frequency for AC Voltage Measurement: 45-65Hz
Voltage Measurement Accuracy: ±(2%rdg+3V)
Current Measurement Range: 0.01nA-8.00mA
Current Measurement Accuracy: ±(2%rdg+0.5nA)
Capacitance Measurement Range: 5.0nF-50.0µF
Capacitance Measurement Accuracy: ±(5%rdg+5nF)
Discharge Rate: From 12kV to 50V within ≤0.47s/µF as specified in the catalog
Insulation Resistance Analysis Functions
IR (Insulation Resistance): Measures electrical insulation resistance
IR(t): Checks insulation resistance over time
DAR (Dielectric Absorption Ratio): Analyzes dielectric absorption ratio
PI (Polarization Index): Analyzes polarization index of insulation
SV (Step Voltage): Insulation testing using multiple voltage levels
Ramp Test: Testing by changing voltage levels according to a specified pattern
Data Display and Recording System
Large color screen for displaying measurement results
Displays real-time changes in insulation resistance, voltage, and current
Saves test results with date and time stamps
Exports data via USB
Power Supply and Operating Environment
AC Power Supply: 85-264VAC, 45-65Hz
Battery Capacity: 5200mAh
Operating Temperature: -20 to 50°C
Operating Humidity: Less than 95% RH without condensation
Protection Level: IP65 when closed / IP40 when open
Examples and Usage Guidelines
Check the insulation resistance of high-voltage power cables
Inspect the condition of transformer insulation
Test the insulation of electric motors and generators
Apply for industrial electrical system inspection and maintenance work
Analyze insulation quality trends using PI, DAR, and IR(t) functions
Record data to compile reports on electrical equipment testing
Limits and Considerations
The maximum test voltage for the K1224 IN model is 12kV, which must be selected appropriately based on the tested device
The highest measurement range of 24TΩ is at a test voltage of 12kV and has an accuracy of ±20% in the range of 2.4TΩ-24TΩ
Do not test insulation while power supply remains connected to the device
Must ensure that the equipment is disconnected from power sources and there are no residual voltages before starting tests
After testing, discharge until safe levels are reached before touching the tested equipment or test leads
Select measurement cables and accessories that support the highest test voltage as per manufacturer specifications
Check the voltage adjustment steps in the 10-12kV range and safety standards of the model before actual use
High-voltage insulation testing should be performed by trained personnel with electrical safety training
Note: Technical data referenced from SCMA General Catalog Vol.5, K1224 IN model supporting a maximum test voltage of 12kV, differing from the K1020 IN model which supports up to 10kV.