ข้อมูลเชิงเทคนิค KEPLER เครื่องทดสอบฉนวนไฟฟ้า (Digital High Voltage Insulation Testers) รุ่น K6213A IN
Test voltage From 500VDC to 10kVDC Adjustable in 500V steps
Preset buttons 1kV / 2.5kV / 5kV / 10kV
Measuring ranges 800kΩ~500GΩ (auto-ranging)
Accuracy ±5% ±2 digits
Output power limit 1W
Voltage regulation Selected voltage +20%-5% of nominal value unless current limited. Meaning that if output current is too high, the voltage will be lowered automatically.
Dimensions 330(L) x 260(W) x 160(D)mm
Weight (battery included) Approx. 3600g
Power source 1.5V "C" x 8 Alkaline batteries
Safety standard EN 61010-1 CAT III 300V, EN 61326-1